Microprocessor Ph Meter
WLS-PH12
WLS-PH12 is a high precision pH meter with large backlit LCD display. The meter can display many useful prompt messages to helps you quickly and easily measuring the samples. pH meter – 5 Point Calibration.
About the product
PH Mode:
Up to 5 points push-button calibration with auto-buffer recognition. Selectable pH buffer Standards, (USA, DIN, NIST, User-defined) o o resolution (0.1, 0.01, 0.001pH) temperature units (C or F) automatically calculates and displays electrode slope that helps you decide whether to replace pH electrode. Calibration due reminder prompts you to calibrate the meter regularly.
ORP Mode:
Absolute or Relative mV measurements provides accurate ORP measured value .1-point offset calibration allows user adjusts displayed value to a known standard.
Features & Benefits
- During the pH mode, up to 5 points push-button calibration with auto- buffer recognition.
- Selectable pH buffer standards and temperature units.
- Automatically calculates and displays electrode slope after calibration.
- Automatic Temp. Compensation (ATC) ensures highly accurate measurement over the entire range.
- Calibration due reminder prompts you to calibrate the meter regularly.
- During ORP mode, Single point offset calibration allows user adjusts displayed value to a known standard.
- Absolute or Relative millivolt (R.mV) measurement provides accurate ORP readings.
- Automatic endpoint function freezes the stable measured values for easy reading and recording data.
- Help message as an operational guide to helps you understand how to use meter.
- Setup menu lets user customizes buffer standards, calibration points, stability conditions, resolution, temperature units date and time to meet measurement requirement.
- Reset function can resumes all settings to factory default values.
- Expanded memory stores and recalls up to 500 readings.
- Built-in real-time clock stamps stored data to meets GLP standards.
- Stored data can be transferred into computer via USB communication interface.
Product Specifications
Model No. | WLS-PH12 | |
PH | pH Range | -2.000 to 20.000pH |
pH Accuracy | ±0.002pH | |
pH Resolutions | 0.001,0.01,0.1 pH (Selectable) | |
Calibrations | Up to 5 Point | |
Calibration due reminder | 1 to 30 days or off | |
pH Buffer Option | USA(pH1.68,4.01,7.00,10.01,12.45), NIST (pH1.68,4.01,6.86,9.18,12.45)DIN (pH1.09,4.65,6.79,9.23,12.7)OR User Define | |
mV | mV Range | ±1999mV |
mV Accuracy | ±0.2mV | |
mV Resolution | 0.1, 1mV (Selectable) | |
Temp. | Range | 0~105°C, 32~212°F |
Accuracy | ±0.5°C, ±0.9°F | |
Temperature compensations | 0~100°C, 32~212°F, Automatic or Manual | |
General | Auto-Off | Manual or Automatic (10, 20, 30 minutes after last key pressed (Selectable) |
Memory | 500 readings with date & time | |
Power Supply | DC 5V adapters, using AC 220V / 50Hz, | |
Dimensions LxWxH | 210×188×60 mm | |
Display | LCD (135x75) mm | |
Power Requirements | DC5V, Using AC adaptor,220c/50hz |